GBS-1000 Programming Manual
28
:CHANnel<X>:OFFSet
Set
Query
Sets or returns the offset level for each channel. The
offset level range is dependent on the vertical scale.
< Long >
< Short >
:channel<X>:offset <NR3>
:channel<X>:offset?
:chan<X>:offs <NR3>
:chan<X>:offs?
<X>
Channel
<NR3>
Offset level
1/2/3/4
CH1/2/3/
4
±0.5
0.5V ~ +0.5V
(2mV/div~50mV/div)
±5.0
5.0V ~ +5.0V
(100mV/div~500mV/div)
±50.0
50.0V ~ +50.0V
(1V/div ~ 5V/div)
:channel1:scale 1.00e2
:channel1:offset 2.00e2
Set CH1 scale to 10mV
Set CH1 offset to 20mV
:CHANnel<X>:PROBe
Set
Query
Sets or returns the probe attenuation factor.
Same as: Channel key F4
< Long >
< Short >
:channel<X>:probe <NR3>
:channel<X>:probe?
:chan<X>:prob <NR1>
:chan<X>:prob?
<X>
Channel
<NR1>
Probe attenuation factor
1/2/3/4
CH1/2/3/4
0
1x
1
10x
2
100x
:channel1:probe 1
Set CH1 probe to 10x