COMMAND DETAILS
23
31B:CHANnel<X>:SCALe
Set
Query
Description
Sets or returns the vertical scale. The scale depends
on the probe attenuation factor.
Same as: Volts/Div knob
Syntax
< Long > < Short >
:channel<X>:scale <NR3>
:channel<X>:scale?
:chan<X>:scal <NR3>
:chan<X>:scal?
Parameter
<X>
Channel
<NR3>
Vertical scale
1/2 CH1/2 2e–3 ~ 5e+0 2mV ~ 5V
(Probe x1)
2e–2 ~ 5e+1 20mV ~ 50V
(Probe x10)
2e–1 ~ 5e+2 200mV ~ 500V
(Probe x100)
Example :channel1:probe 0
:channel1:scale 2.00e–3
Sets the Channel 1 probe
attenuation factor to x1
Sets the Channel 1
vertical scale to 2mV/div
GDS-1000 Programming Manual
24
9BCursor Command
:CURSor:X<X>Position................................................214H24
:CURSor:Y<X>Position ................................................215H25
:CURSor:<X>DELta......................................................216H25
:CURSor:<X>DISplay...................................................217H26
:CURSor:SOURce.........................................................218H27
32B:CURSor:X<X>Position
Set
Query
Description
Sets or returns the horizontal (X axis) cursor
position.
Same as: Cursor key → F5 (X-Y) →F2 (X1) or F3
(X2) + Variable knob
Syntax
< Long > < Short >
:cursor:x<X>position <NR1>
:cursor:x<X>position?
:curs:x<X>p <NR1>
:curs:x<X>p?
Parameter
<X>
Cursor 1 or 2
<NR1>
Cursor position
1 Cursor X1 1 ~ 249 1 ~ 249 point
2 Cursor X2
Note
When in the query mode, the returned data format
is <NR3> as follows.
CH1, CH2, Math (CH1±CH2): time (s)
Math (FFT): frequency (Hz)
Example :cursor:xdisplay 1
:cursor:x1position 100
Puts the horizontal
cursor X1 on the 100
point position
:channel:math 3
:cursor:xdisplay 1
:cursor:x1position?
→ 2.500E+03
Returns the X1 cursor
position as 2500Hz in
the Math FFT mode