GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
29
V/I:
Test Function Unit (2 bytes)
ACW KV
DCW KV
IR V
Cnt (GB) A
CAC KV
CDC KV
HOLD PAUSE
Meas:
Test Function Unit (4 bytes)
ACW mA
DCW mA
IR MOhm
Cnt (GB) mOhm
CAC mA
CDC mA
HOLD
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
30
Time:
Time (1 byte) Value (5 bytes) Unit (1 bytes)
Ramp 005.0 sec
Test 005.0 sec
Examples:
User the command :MEAS3? after measurement
Display for PC 10,1,0.100kV,01.00mA, T005.0S
10=Pass
1=ACW
0.100kV=Mesuremett voltage 100V
01.00mA=Mesurement current 1mA
T005.0S=Test time 5 sec
:UTILity:AC:FREQuency
Function:
Set or query the ACW CAC frequency ( 0:50Hz, 1:60Hz )
Syntax:
:UTILity: FREQuency <Boolean>
:UTILity: FREQuency?
50 (50Hz) or 60 (60Hz)
Return:
50 | 60