Gw-instek GOM-802 Programming Manual Manual de usuario Pagina 19

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 27
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 18
DC MILLI-OHM METER
PROGRAMMER MANUAL
36
:CALCulation:LIMit:LOWer
Function:
Returns or sets the value of the lower limit in the compare mode.
Syntax:
:CALCulation:LIMit:LOWer?
Returns: <NR2>
<NR2> is in the range from 0 to 99.9.
:CALCulation:LIMit:LOWer <NR2>
Examples:
:CALCulation:LIMit:LOWer? might return “+10.0 ” in the COMP
mode. It sets down limit percentage to 10%.
:CALCulation:LIMit:LOWer 20 sets 20% down limit percentage in the
COMP mode.
:CALCulation:LIMit:UPPer
Function:
Returns or sets the value of the upper limit in the compare mode.
Syntax:
:CALCulation:LIMit:UPPer?
Returns: <NR2>
<NR2> is in the range from 0 to 99.9.
:CALCulation:LIMit:UPPer <NR2>
DC MILLI-OHM METER
PROGRAMMER MANUAL
37
Examples:
:CALCulation:LIMit:UPPer? might return “+10.0 ” in the COMP mode.
It sets up limit percentage to 10%.
:CALCulation:LIMit:UPPer 20 sets 20% up limit percentage in the
COMP mode.
:CALCulation:LIMit:FAIL?(query only)
Function:
Returns limit result.
Syntax:
:CALCulation:LIMit:FAIL?
Returns:
0 (Low)/1 (Pass)/2 (High)
Examples:
:CALCulation:LIMit:FAIL? might return “1” to indicate pass.
:CALCulation:LIMit:NORMal
Function:
Returns or sets the normal value.
Syntax:
:CALCulation:LIMit:NORMal?
Returns: <NR1>
<NR1> is in the range from 0 to 33000.
Vista de pagina 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 25 26 27

Comentarios a estos manuales

Sin comentarios