GDS-2000 Series Programming Manual
23
Combining Commands
You can use a semicolon (;) to combine commands and queries. This oscilloscope
executes coherent commands in the order it receives them. When you coherent
queries, the oscilloscope will combine the responses into a single response
message with LineFeed (0x0A) for each command. For example, if the frequency
and amplitude of the signal are equal to 1kHz and 1V, the command
:MEASure:FREQuency?;:MEASure:VAMPlitude?
returns the message
1.000E+3 1.000E+0
*: Please note that a LineFeed (0x0A) message terminator will be added behind each responded
command.
Therefore the return message of “1.000E+3 1.000E+0” should be exact looked like as following
example in HEX mode: