Gw-instek GDS-2000 series Programming manual Manual de usuario Pagina 37

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 93
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 36
GDS-2000 Series Programming Manual
37
:CHANnel<X>:SCALe
Sets or query the vertical scale of the specified channel.
Syntax
:CHANnel<X>:SCALe <NR3>
:CHANnel<X>:SCALe?
Arguments
<X>Specify the channel number (1|2|3|4)
<NR3> is the desired gain value in volts per division. The range is 2mV/div to 5V/div
(with 1X probe).
Next table shows the relationship between the <NR3> value and matching scale.
2.00E-3Æ2mV 1.00E-2Æ10mV 1.00E-1Æ100mV 1.00E+0Æ1V
5.00E-3Æ5mV 2.00E-2Æ20mV 2.00E-1Æ200mV 2.00E+0Æ2V
5.00E-2Æ50mV 5.00E-1Æ500mV 5.00E+0Æ5V
Returns
<NR3>
Examples
:CHANnel4:SCALe 1.00E+0, setup the channel four at 1V per division.