GDS-2000 Series Programming Manual
47
:GONogo:VIOLation
Sets and query the user-defined GO/NO-GO violation conditions. This command is
equal to the function key “F3” pressing of GO/NO-GO menu.
Syntax
:GONogo:VIOLation {0|1|2|3}
:GONogo:VIOLation?
Arguments
0→Stop 1→Stop + beep
2→Continue 3→Continue + beep
Returns
<NR1>
Note: Before running any Go/NoGo commands, the oscilloscope has to be initialized, please
add the command of “:GONogo:FUNCtion 1” in order to initial the oscilloscope.