:MEASure:SCAN
Channel Specific
Command
Description
This command allows the mainframe to scan all
the load modules.
Syntax
:MEASure:SCAN {OFF|0|ON|1}
Parameter
OFF/0
Disabled
ON/1
Enabled
Example
:MEAS:SCAN 0
Disable scanning.
Query Syntax
:MEASure:SCAN? <NR1>
Return Parameter
<NR1>
Scan
0
Disabled
1
Enabled
Query Example
:MEAS:SCAN?
1
Returns the scanning
status. Here scanning is
enabled.
:MEASure:ALLVoltage?
All Channel
Command
Description
This query measures the voltage values of all the
load modules/channels in order from 1-8 (PEL-
2004)/1-4(PEL2002).
Query Syntax
:MEASure:ALLVoltage? <aard>
Query Parameter
<aard> 1 unit = 1 volt
CH1,CH2,CH3,CH4,CH5,
CH6,CH7,CH8
Returns all the voltage
values from all the
channels, 1-8(PEL-
2004)/1-4(PEL-2002).