37
Example
:measure:source 1
:measure:vhi?
Selects Channel 1, and
then measures the
global high Voltage.
:MEASure:VLO
Query
Description
Returns the global low voltage.
Same as: Measure key F1~F5 F3 (Vlo)
Syntax
< Long >
< Short >
:measure:vlo?
:meas:vlo?
Returns
<NR3>
Note
Before using this command, select the
measurement channel. See the example below.
Example
:measure:source 1
:measure:vlo?
Selects Channel 1, and
then measures the
global low Voltage.
:MEASure:VMAX
Query
Description
Returns the maximum amplitude.
Same as: Measure key F1~F5 F3 (Vmax)
Syntax
< Long >
< Short >
:measure:vmax?
:meas:vmax?
Returns
<NR3>
Note
Before using this command, select the
measurement channel. See the example below.
Example
:measure:source 1
:measure:vmax?
Selects Channel 1, and
then measures the
maximum amplitude.
Vista de pagina 36