Gw-instek GDS-1000-U Series Programming Manual Manual de usuario Pagina 50

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 70
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 49
GDS-1000-U Programming Manual
50
Data Log Commands
:DATALOG:STATE ....................................................... 50
:DATALOG:SOURce .................................................... 50
:DATALOG:SAVe ......................................................... 51
:DATALOG:INTerval .................................................... 51
:DATALOG:DURation .................................................. 51
:DATALOG:STATE
Set
Query
Description
Turns the datalogging function on/off.
Same as: Utility key More (F5) Data Logging
Menu(F3)Data Logging (F1).
Syntax
< Long >
< Short >
:DATALOG:STATE {0|1}
:DATALOG:STATE?
:DATALOG:STATE
{0|1}
:DATALOG:STATE?
Parameter/
Return parameter
0
Off. Stop data logging function.
1
On. Start data logging function.
Example
:DATALOG:STATE 1
Turn data logging on.
:DATALOG:SOURce
Set
Query
Description
Sets or queries the data logging source channel.
Syntax
< Long >
< Short >
:DATALOG:SOURce{1|2}
:DATALOG:SOURce?
:DATALOG:SOUR{1|2}
:DATALOG:SOUR?
Parameter/
Return parameter
1
Sets CH1 as the source channel
2
Sets CH2 as the source channel
Example
:DATALOG:SOUR 1
Set source as CH1.
Vista de pagina 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 69 70

Comentarios a estos manuales

Sin comentarios