COMMAND DETAILS
41
:MEASure:RPReshoot
Query
Description
Returns rise overshoot amplitude in percentage.
Same as: Measure key → F1~F5 → F3 (RPReshoot)
Syntax
< Long > < Short >
:measure:rpreshoot? :meas:rpr?
Returns <NR2> with % sign
Note
Before using this command, select the
measurement channel. See the example below.
Example :measure:source 1
:measure:rpreshoot?
Selects Channel 1, and
then measures the rise
preshoot.
:MEASure:SOURce
Set
Query
Description
Selects the measurement channel.
Same as: Measure key → F1~F5 → F1, F2
Syntax
< Long > < Short >
:measure:source <NR1>
:measure:source?
:meas:sour <NR1>
:meas:sour?
Parameter
<NR1>
1 ~ 2 Channel1 ~ 2
Example :measure:source 1
:measure:rprshoot?
Selects Channel 1, and
then measures the rise
preshoot.
:MEASure:VAMPlitude
Query
Description
Returns the voltage difference between positive
and negative peak.
Same as: Measure key → F1~F5 → F3 (Vamp)
GDS-1000A Programming Manual
42
Syntax
< Long > < Short >
:measure:vamplitude? :meas:vamp?
Returns <NR3>
Note
Before using this command, select the
measurement channel. See the example below.
Example :measure:source 1
:measure:vamplitude?
Selects Channel 1, and
then measures the rise
Voltage amplitude.
:MEASure:VAVerage
Query
Description
Returns the average voltage.
Same as: Measure key → F1~F5 → F3 (Vavg)
Syntax
< Long > < Short >
:measure:vaverage? :meas:vavg?
Returns <NR3>
Note
Before using this command, select the
measurement channel. See the example below.
Example :measure:source 1
:measure:vaverage?
Selects Channel 1, and
then measures the
average Voltage.
:MEASure:VHI
Query
Description
Returns the global high voltage.
Same as: Measure key → F1~F5 → F3 (Vhi)
Syntax
< Long > < Short >
:measure:vhi? :meas:vhi?
Returns <NR3>
Note
Before using this command, select the
measurement channel. See the example below.
Comentarios a estos manuales