Gw-instek GDS-3000 Series Programming Manual Manual de usuario Pagina 49

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 103
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 48
COMMAND DETAILS
49
Example
:CURSor:H2Position 1.5E-3
Sets the H2 cursor position to 1.5ms.
:CURSor:HDELta
Query
Description
Returns the delta of H1 and H2.
Syntax
:CURSor:HDELta {?}
Return Parameter
<NR3>
Returns the distance between two
horizontal cursors.
Example
:CURSor:HDELta?
5.0E-9
Returns the horizontal delta as 5ns.
:CURSor:V1Position
Set
Query
Description
Sets the first vertical cursor (V1) position.
Syntax
:CURSor:V1Position {<NRf>| ?}
Parameter
<NRf>
Position
Depends on the vertical scale
Return parameter
Returns the cursor position.
Example
:CURSor:V1Position 1.6E -1
Sets the V1 cursor position to 160mA.
:CURSor:V2Position
Set
Query
Description
Sets the first vertical cursor (V2) position.
Syntax
:CURSor:V2Position {<NRf> | ?}
Parameter
<NRf>
Position
Depends on the vertical scale
Return parameter
Returns the cursor position.
Vista de pagina 48
1 2 ... 44 45 46 47 48 49 50 51 52 53 54 ... 102 103

Comentarios a estos manuales

Sin comentarios