COMMAND DETAILS
83
Returns the status of the frequency filter.
:TRIGger:REJect OFF
Turns the frequency filter off.
Sets or queries the trigger mode.
:TRIGger:MODe {AUTo | NORMal | ?}
Auto trigger (Untriggered roll)
Returns the trigger mode.
:TRIGger:MODe NORMal
Sets the trigger mode to normal.
Sets or queries the holdoff time.
:TRIGger:HOLDoff {<NRf> | ?}
Returns the trigger holdoff time as <NR3>.
:TRIGger:HOLDoff 1.00E-8
Sets the trigger holdoff time to 10ns.
Sets or queries the high trigger level (applicable for
Rise and Fall/Pulse Runt trigger).