Measure Commands
MEASure[:SCALar]:CURRent[:DC] ................................... 59
MEASure[:SCALar]:VOLTage[:DC] .................................. 59
MEASure[:SCALar]:POWer[:DC] ....................................... 59
MEASure[:SCALar]:CURRent[:DC]
Query
Description
Takes a measurement and returns the average
output current
Syntax
MEASure[:SCALar]:CURRent[:DC]?
Return parameter
<NRf>
Returns the current in amps.
MEASure[:SCALar]:VOLTage[:DC]
Query
Description
Takes a measurement and returns the average
output voltage.
Syntax
MEASure[:SCALar]:VOLTage[:DC]?
Return
<NRf>
Returns the voltage in volts.
MEASure[:SCALar]:POWer[:DC]
Query
Description
Takes a measurement and returns the average
output power.
Syntax
MEASure[:SCALar]:POWer[:DC]?
Return
<NRf>
Returns the power measured in watts.