PSW Series Programming Manual
88
Parameter/
Return parameter
SYST:COMM:USB:REAR:MODE 1
Sets the rear panel USB-B port mode to GPIB-USB
adapter.
Queries the error queue. The last error message is
returned. A maximum of 32 errors are stored in the
error queue.
Returns an error code followed by
an error message as a string. The
string is returned as “string”.
SYSTem:ERRor?
-100, “Command error”
Sets or queries the key lock mode. This setting is
the equivalent of the F-19 function setting.
SYSTem:KEYLock:MODE {0|1}
SYSTem:KEYLock:MODE?
Parameter /
Return parameter
Panel lock: allow output off.
Panel lock: allow output on/off.
Enables or disables the front panel key lock.
SYSTem:KLOCk { OFF | ON | 0 | 1}
SYSTem:KLOCk?