REMOTE CONTROL
103
Parameter
<NR1>
0~255
Return parameter
<NR1>
Returns the bit sum of the Service Request
Enable register.
*STB
Query
Description
Queries the bit sum of the Status Byte register with
MSS (Master summary Status) replacing the RQS
bit (bit 6).
Query Syntax
*STB?
Return parameter
<NR1>
Returns the bit sum of the Status Byte register
with the MSS bit (bit 6).
*TRG
Set
Description
The *TRG command is able to generate a “get”
(Group Execute Trigger). If the PSU cannot accept
a trigger at the time of the command, an error
message is generated (-211, “Trigger ignored”).
Syntax
*TRG
*TST
Query
Description
Executes a self test.
Query Syntax
*TST?
Return parameter
0
Returns “0” if there are no errors.
<NR1>
Returns an error code <NR1> if there is an
error.
*WAI
Set
Description
Prevents any other commands or queries from
being executed until all outstanding commands
have completed.
Syntax
*WAI