REMOTE CONTROL
79
Query Syntax
[:SOURce]:CURRent:PROTection:DELay?
Parameter
<NR2>
0.1~2.0 seconds, where 0=no delay
MAX
The maximum allowed delay time
MIN
The minimum allowed delay time
Return parameter
<NR2>
Returns the delay time in seconds
Example
SOUR:CURR:PROT:DEL MAX
Sets the current protection delay to the maximum.
[:SOURce]:CURRent:PROTection[:LEVel]
Set
Query
Description
Sets or queries the OCP (over-current protection)
level in amps.
Syntax
[:SOURce]:CURRent:PROTection[:LEVel] {<NR2>(A)
|MINimum|MAXimum}
Query Syntax
[:SOURce]:CURRent:PROTection[:LEVel]?
Parameter
<NR2>
Current protection level.
Minimum: Depend on the unit type:
if Irated * 0.1 > 5A, then minimum = 5A,
else minimum = Irated * 0.1
Maximum: Irated * 1.1
MIN
Minimum current level.
MAX
Maximum current level.
Return parameter
<NR2>
Returns the current protection level.
Example
SOUR:CURR:PROT:LEV?
+5.000
Returns the minimum possible current level in amps.
[:SOURce]:CURRent:PROTection:STATe
Set
Query
Description
Turns OCP (over-current protection) on or off.
Syntax
[:SOURce]:CURRent:PROTection:STATe
{<bool>|OFF|ON}
Query Syntax
[:SOURce]:CURRent:PROTection:STATe?