Gw-instek PSU-Series Programming User Manual Manual de usuario Pagina 85

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 128
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 84
REMOTE CONTROL
85
[:SOURce]:VOLTage[0]:PROTection[:LEVel]
Set
Query
Description
Sets or queries the overvoltage protection level.
Syntax
[:SOURce]:VOLTage[0]:PROTection[:LEVel]
{<NR2>(V)|MINimum|MAXimum}
Query Syntax
[:SOURce]:VOLTage[0]:PROTection[:LEVel]?
Parameter/Return
<NR2>
Minimum: Depends on the unit type:
if Vrated * 0.1 > 5V, then Minimum = 5V,
else Minimum = Vrated * 0.1
Maximum: Vrated * 1.1
MIN
Minimum OVP level
MAX
Maximum OVP level
Example
SOUR:VOLT:PROT:LEV MAX
Sets the OVP level to its maximum.
[:SOURce]:VOLTage[0]:PROTection:TRIPped
Set
Query
Description
Sets or queries the overvoltage protection level.
Query Syntax
[:SOURce]:VOLTage[0]:PROTection:TRIPped?
Return parameter
<bool>
0
Protection not tripped
1
Protection tripped
Example
SOUR:VOLT:PROT:TRIP?
>0
Indicates that the OVP protection has not been
tripped.