COMMAND DETAILS
125
BACKward
Moves to the previous waveform in
memory
<NRf>
Sets the current waveform number.
The waveform number must be equal
to or less than the total number of
waveforms, as returned by the
REPLAY:TOTalnum? query.
Example
REPLAY:TOTalnum?
8.960E+02
REPLAY:CURRent 100
REPLAY:TOTalnum?
1.000E+02
Indicates that there are a total of 896 waveforms.
Set the current waveform to #100. The return value
(100) indicates that the current waveform is #100.
Rotate Commands
The rotate function sets the screen orientation.
:ROTATE ....................................................................125
:ROTATE
Description
Sets or queries the screen orientation.
Syntax
:ROTATE { PORTrait | LANDscape | 0 | 1 | ? }
Parameter/Return
parameter
PORTrait / 0
PORTrait
LANDscape / 1
LANDscape
Set
Query