COMMAND DETAILS
57
Hardcopy Commands
:HARDcopy:START ..................................................... 57
:HARDcopy:ASSIGN ................................................... 57
:HARDcopy:START
Description
Executing the HARDcopy:START command is the
equivalent of pressing the Save hardcopy key on
the screen.
Syntax
:HARDcopy:START
Related
Commands
:HARDcopy:ASSIGN
:HARDcopy:ASSIGN
Description
Sets or queries what file type the Save hardcopy
key has been assigned to save.
Syntax
:HARDcopy:ASSIGN
{IMAGe | WAVEform | SETUp | ALL | ?}
Related
Commands
:HARDcopy:START
Parameter
IMAGe
Save image files.
WAVEform
Save waveforms.
SETUp
Save the panel setup.
ALL
Save All (image, waveform,setup)
Return parameter
Returns the file type.
(IMAGE/WAVEFORM/SETUP/ALL)
Example
:HARDcopy:ASSIGN IMAGE.
Set the hardcopy key to save image files.
Set
Set
Query