GDS-200 & GDS-300 Programming Manual
48
Return parameter
<NR3>
Returns the cursor position.
Example
:CURSor:V2Position 1.1E-1
Sets the V2 cursor position to 110mV.
:CURSor:VDELta
Description
Returns the delta of V1 and V2.
Syntax
:CURSor:VDELta?
Return Parameter
<NR3>
Returns the difference between two
vertical cursors.
Example
:CURSor:VDELta?
4.00E+0
Returns the vertical delta as 4 volts.
:CURSor:XY:RECTangular:X:POSition<X>
Description
Sets or queries the horizontal position in XY mode
for the X rectangular coordinates for cursor 1 or 2.
Syntax
:CURSor:XY:RECTangular:X:POSition<X> {NRf|?}
Parameter
<X>
Cursor 1, 2
<NRf>
Horizontal position co-ordinates
Return parameter
<NR3>
Returns the cursor position.
Example
:CURSor:XY:RECTangular:X:POSition1 4.0E-3
Sets the X-coordinate cursor 1 position to
40mV/mV.
:CURSor:XY:RECTangular:X:DELta
Description
Returns the delta value of cursor 1 and 2 on the X
coordinate.
Syntax
:CURSor:XY:RECTangular:X:DELta?
Query
Set
Query
Query