GDS-806/810/820/840 Programming Manual
52
:TEMPlate:MIN (Available for firmware 2.0 and above only)
Sets and query the number of template used for minimum boundary.
Before running this command, the oscilloscope has to be initialized as “Normal
Template” mode, please add the command of “:TEMPlate:MODe 0 in order to
change the settings.
Syntax
:TEMPlate:MIN {1|100}
:TEMPlate:MIN?
Returns
<NR1>
Note: A template can be defined as either MAX or MIN template only.
Note B: Before running this command, the oscilloscope has to be initialized, please add the
command of “:GONogo:FUNCtion 1 in order to initial the oscilloscope.
:TEMPlate:MODe (Available for firmware 2.0 and above only)
Sets and query the template mode. When we select Auto mode, the auto template
source is coming from CH1 or CH2. When we set normal mode, the template is
coming from 1 of the 100 templates or RefA or RefB.
Syntax
:TEMPlate:MODe {0|1}
:TEMPlate:MODe?
Arguments
0 Select normal template. 1 Select auto template.
Returns
<NR1>