GDS-806/810/820/840 Programming Manual
76
Status Registers.
There are two status registers are included with this oscilloscope which defined by
IEEE-488.1 and IEEE-488.2 standards.
z Status Byte Register (SBR)
z Standard Event Status Register (SESR)
Status Byte Register (SBR): The SBR (Table 5-3) summaries the status of all other
registers and queue.
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
OPER RQS ESB MAV QUES NU NU NU
128 64 32 16 8 4 2 1
Table 5-3: Status Byte Register (SBR)
From the bit 0 to bit 2 are not used, these bits are always zero.
The bit 3 (QUES, QUEStionable) is the summary bit for the QESR (QUEStionable
Event Status Register). When this bit is high it indicates that status is enabled and
present in the QESR.
The bit 4 (MAV, Message Available) indicates that output is available in the output
queue.
The bit 5 (ESB, Event Status Bit) is the summary bit for the Standard Event Status
Register (SESR). When this bit is high it indicates that status is enabled and
present in the SESR.
The bit 6 (RQS, Request Service) is obtained from a serial poll. This bit shows that
this oscilloscope requests service from the GPIB controller.
The bit 7 (OPER, OPERation) is the summary bit for the OESR (OPERation EVENt
STATus Register).