Gw-instek GDS-2000A series Programming Manual Manual de usuario Pagina 135

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 272
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 134
COMMAND DETAILS
135
SPI
SPI mode
UART
UART mode
PARALLEL
Parallel mode
Example
:TRIGger:BUS:TYPe?
UART
:TRIGger:BUS:B1:I2C:CONDition
Description
Sets or queries the I
2
C trigger conditions.
Syntax
:TRIGger:BUS:B1:I2C:CONDition
{STARt | STOP | REPEATstart | ACKMISS | ADDRess |
DATA | ADDRANDDATA | ? }
Parameter
STARt
Set Start as the I
2
C trigger
condition.
STOP
Set Stop as the I
2
C trigger
condition.
REPEATstart
Set Repeat of Start as the I
2
C
trigger condition.
ACKMISS
Set Missing Acknowledgement
as the I
2
C trigger condition.
ADDRess
Set Address as the I
2
C trigger
condition.
DATA
Set Data as the I
2
C trigger
condition.
ADDRANDDATA
Set Address and Data as the I
2
C
trigger condition.
Return parameter
Returns the I
2
C bus trigger condition.
Example
:TRIGger:BUS:B1:I2C:CONDition ADDRess
Set Address as the I2C trigger condition.
Set
Query
Vista de pagina 134
1 2 ... 130 131 132 133 134 135 136 137 138 139 140 ... 271 272

Comentarios a estos manuales

Sin comentarios