Gw-instek GDS-2000A series Programming Manual Manual de usuario Pagina 51

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 272
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 50
COMMAND DETAILS
51
Return Parameter
ON
Channel is on.
OFF
Channel is off
Example
:CHANnel1:DISPlay ON
Turns on Channel 1
:CHANnel<X>:EXPand
Description
Sets Expand By Ground or Expand By Center for a
channel or queries its status.
Syntax
:CHANnel<X>:EXPand {GND | CENTer | ?}
Parameter
<X>
Channel 1,2,3,4
GND
Ground
CENTer
Center
Return parameter
GND
Expand By Ground
CENTER
Expand By Center
Example
:CHANnel1:EXPand GND
Sets Channel 1 to Expand By Ground.
:CHANnel<X>:IMPedance?
Description
Returns the impedance of the oscilloscope.
Syntax
:CHANnel<X>:IMPedance?
Parameter
<x>
Channel
1/2/3/4
CH1/2/3/4
Return parameter
<NR3>
Returns the impedance value.
Example
:CHANnel1:IMPedance?
1.000000E+06
The impedance is 1M ohms.
Set
Query
Query
Vista de pagina 50
1 2 ... 46 47 48 49 50 51 52 53 54 55 56 ... 271 272

Comentarios a estos manuales

Sin comentarios