Gw-instek GDS-2000A series Programming Manual Manual de usuario Pagina 149

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 272
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 148
COMMAND DETAILS
149
Example1
:TRIGger:LOGic:PATtern:WHEn FALSE
Set the logic to false.
Example2
:TRIGger:LOGic:PATtern:WHEn?
FALSE
:TRIGger:TIMEOut:WHEn
Description
Sets or queries the timeout trigger condition.
Syntax
:TRIGger:TIMEOut:WHEn {HIGH|LOW|EITher|?}
Related
commands
:TRIGger:TIMEOut:TIMER
Parameter
HIGH
Signal is high.
LOW
Signal is low.
EITher
Signal is high or low.
Return parameter
Returns the timeout condition (HIGH, LOW, EITHER).
Example1
:TRIGger:TIMEOut:WHEn LOW
Sets the timeout condition to low.
:TRIGger:TIMEOut:TIMER
Description
Sets or returns timeout trigger time.
Syntax
:TRIGger:TIMEOut:TIMER {value | ? }
Related
commands
:TRIGger:TIMEOut:WHEn
Parameter/Return
parameter
<value>
<NR3> Timeout time. (10nS to 10S).
Example
:TRIGger:TIMEOut:TIMER?
8.960e-05
Set
Query
Set
Query