Vista de pagina 127
GSP-930 Programming Manual
126
<integer>
<NR1>
:AVER:COUN 20
[:SENSe]:AVERage:STATe
Set
Query
Turns the Average function on/off or queries its
state.
[:SENSe]:AVERage:STATe {OFF|ON|0|1}
[:SENSe]:AVERage:STATe?
0
Turn the Average function off.
1
Turn the Average function on.
OFF
Turn the Average function off.
ON
Turn the Average function on.
0
The Average function is off.
1
The Average function is on.
:AVER:STAT ON
[:SENSe]:AVERage:TYPE
Set
Sets the method that the Average function uses to
calculate the average.
[:SENSe]:AVERage:TYPE
{VOLTage|LOGarithmic|POWer}
[:SENSe]:AVERage:TYPE?
VOLTage
Sets Average to voltage
LOGarithmic
Sets Average to logarithmic
POWer
Sets Average to power
:AVER:TYPE VOLT
[:SENSe]:BANDwidth|BWIDth[:RESolution]
Set
Query
Sets or queries the resolution bandwidth (RBW).
Vista de pagina 127