Gw-instek GSP-930 Programming Manual Manual de usuario Pagina 74

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 195
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 73
GSP-930 Programming Manual
72
:CALCulate:DEMod:AM:RESult:MAXimum?
Query
Returns the maximum recorded measurement
results for AM demodulation as a comma
separated string.
:CALCulate:DEMod:AM:RESult:MAXimum?
<depth,rate,power,offset,sinad>
depth
Modulation depth in %.
rate
Modulation rate in Hz.
power
Carrier power in the current Y-axis units.
offset
Carrier frequency offset in Hz.
sinad
Signal to noise and distortion ratio in dB.
:CALC:DEM:AM:RES:MAX?
>9.840e+1,1.02e+2,-1.12e+1,3.21e+1,1.61e+1
:CALCulate:DEMod:AM:STATe
Set
Query
Sets or queries the state of the AM Analysis
function.
:CALCulate:DEMod:AM:STATe {ON|OFF|1|0}
:CALCulate:DEMod:AM:STATe?
0
Turn AM Analysis off.
1
Turn AM Analysis on.
OFF
Turn AM Analysis off.
ON
Turn AM Analysis on.
0
AM Analysis is off.
1
AM Analysis is on.
:CALC:DEM:AM:STAT 1
Vista de pagina 73
1 2 ... 69 70 71 72 73 74 75 76 77 78 79 ... 194 195

Comentarios a estos manuales

Sin comentarios