REMOTE CONTROL
139
Parameter
<freq>
<NRf>
Return parameter
<NR3>
Hz
Example
:OCBW:BAND 4.5+6
[:SENSe]:OCBW:PERCent
Set
Query
Description
Sets or queries the OCBW percentage (OCBW %)
parameter.
Syntax
[:SENSe]:OCBW:PERCent <integer>
Query Syntax
[:SENSe]:OCBW:PERCent?
Parameter/
Return parameter
<integer>
<NR1>0~100
Example
:OCBW:PERC 90
[:SENSe]:OCBW:SPACe
Set
Query
Description
Sets or queries the OCBW channel space for
OCBW measurements.
Syntax
[:SENSe]:OCBW:SPACe <freq>
Query Syntax
[:SENSe]:OCBW:SPACe?
Parameter
<freq>
<NRf>
Return parameter
<NR3>
Example
:OCBW:SPAC 6e+7
[:SENSe]:PMETer:FREQuency
Set
Query
Description
Sets or queries the power meter measurement
frequency.
Syntax
[:SENSe]:PMETer:FREQuency <freq>
Query Syntax
[:SENSe]:PMETer:FREQuency?