Gw-instek GSP-930 Programming Manual Manual de usuario Pagina 67

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 195
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 66
REMOTE CONTROL
65
:CALCulate:ACPR:ACHannel<n>:LOWer?
Query
Description
Returns the ACPR (adjacent channel power ratio in
dB) calculated for the selected lower adjacent
channel.
Query Syntax
:CALCulate:ACPR:ACHannel<n>:LOWer?
Parameter
<n>
<NR1>adjacent channel 1~3
Return parameter
<NR3>
Power ratio in dB
Query Example
:CALC:ACPR:ACH1:LOW?
>1.801e+01
:CALCulate:ACPR:ACHannel<n>:UPPer?
Query
Description
Returns the ACPR (adjacent channel power ratio in
dB) calculated for the selected higher adjacent
channel.
Query Syntax
:CALCulate:ACPR:ACHannel<n>:UPPer?
Parameter
<n>
<NR1>adjacent channel 1~3
Return parameter
<NR3>
Power ratio in dB
Query Example
:CALC:ACPR:ACH1:UPP?
>1.921e+01
:CALCulate:ACPR:ACHannel<n>:STATe
Set
Query
Description
Sets or queries the state of the selected adjacent
channel.
Syntax
:CALCulate:ACPR:ACHannel<n>:STATe {OFF|ON|0|1}
Query Syntax
:CALCulate:ACPR:ACHannel<n>:STATe?
Parameter
<n>
<NR1>adjacent channel 1~3
0
Disable the selected channel.
1
Enable the selected channel.
OFF
Disable the selected channel.
Vista de pagina 66