COMMNAND OVERVIEW
153
CSummary Status Registers
Overview
The CSummary Status registers indicate the
current operation mode of the load generator.
CSummary Status Registers
4
2
2
1
1
0
CV CR CP
9-15
0
Bit weight
Bit Position
Condition
CV CR CP0PTR/NTR
CV CR CP0EVENT
CV CR CP0ENABLE
PRUN 0 CP
PRUN 0 CP
PRUN 0 CP
PRUN 0 CP
8 4-7 3
256 16 8
Event Bits
CP: The unit has switched to CP mode.
CR: The unit has switched to CR mode.
CV: The unit has switched to CV mode.
CP: The unit has switched to CP mode.
PRUN: A program is running.
Condition
Register
The condition register indicates the status of the
electronic load. The condition register can only
be changed by a change in the condition of the
electronic load. Reading the condition register
does not change the state of the condition
register.
PTR/NTR Filters
The PTR/NTR (Positive/Negative transition)
register determines the type of transition
conditions that will trigger an event. Only the
CSummary Status Register, Questionable Status
Register and Operation Status Register can be
transition programmed.
Event Register
The Event Register indicates if an event has
been triggered according to the transition
settings from the PTR/NTR Register.