COMMNAND OVERVIEW
51
Return parameter
<NR2>
Returns the delay time in
seconds.
Example
:GNG:DTIMe 0.5
Sets the delay time to 0.5 seconds.
Query example
:GNG:DTIMe?
>0.5
The delay time is 0.5 seconds.
[:CONFigure]:GNG:MODE
Set
Query
Description
Sets or queries the entry mode for the Go-NoGo
settings. The entry mode determines whether the
Go-NoGo limits are set as values or as a
percentage value from a center reference value.
Syntax
[:CONFigure]:GNG:MODE {PERCent | VALue }
Query Syntax
[:CONFigure]:GNG:MODE?
Parameter
PERCENT or PERC
Sets the entry mode to %.
VALUE or VAL
Sets the entry mode to
value
Return parameter
Percent
% entry mode
Value
Value entry mode
Example
:GNG:MODE PERCent
Sets the entry mode to %.
Query example
:GNG:MODE?
>Percent
The entry mode is %.
[:CONFigure]:GNG[:PASS]
Set
Query
Description
Queries the Go-NoGo test result(s). This command
can be used for all test modes (CC, CV, CR, CP).
Query Syntax
[:CONFigure]:GNG[:PASS]?
Return parameter
NG
No Go (fail)
GO
Go (Pass)