COMMNAND OVERVIEW
45
Sets or queries the Von voltage settings.
[:CONFigure]:VON {<NRf>[ V ] | MINimum|
MAXimum | LON | LOFF }
The Von voltage level
(default unit is V)
Minimum Von voltage
level
Maximum Von voltage
level
Returns the Von latch
settings.
:VON 10.0V
Sets the Von voltage to 10.0 volts.
:VON?
>Latch OFF, 10.000
The Von voltage level is 10V.
Sets or queries the Von Delay settings in seconds.
[:CONFigure]:VDELay {<NRf>[s] | MINimum |
MAXimum | OFF }
The delay time in seconds
Returns the delay time in
seconds
Comentarios a estos manuales