Gw-instek PEL-3000 Series Programming Manual Manual de usuario Pagina 45

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 174
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 44
COMMNAND OVERVIEW
43
Example1
:OPP LIMIT
Sets the OCP setting to limit.
Example2
:OPP 10.000
Sets the OPP value to 10W.
Query example
:OPP?
>LIMIT, 10.000
The OPP setting is LIMIT and the OPP value is
10.000W.
[:CONFigure]:UVP
Set
Query
Description
Sets or queries the UVP trip settings. The UVP can
also be cleared with this command.
Syntax
[:CONFigure]:UVP {<NRf>[ V ] |MINimum |
MAXimum }
Query Syntax
[:CONFigure]:UVP?
Parameter
<NRf>[V]
voltage limit value.
MINIMUM or MIN
MINIMUM value.
MAXIMUM or MAX
MAXIMUM value.
Return parameter
Returns the UVP level (<NR2>)
Example1
:UVP 10.00
Sets the UVP setting to 10V.
Query example
:UVP?
> 10.0000
The UVP setting is 10.0000V.
[:CONFigure]:OVP
Set
Query
Description
Sets or queries the OVP trip settings. The OVP can
also be cleared with this command.
Syntax
[:CONFigure]:OVP {<NRf>[ V ] | MINimum|
MAXimum}
Vista de pagina 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 173 174

Comentarios a estos manuales

Sin comentarios