COMMNAND OVERVIEW
61
[:CONFigure]:STEP:CVH
Set
Query
Description
Configures the step resolution for CV High Range.
Note: The step resolution setting will be
automatically rounded to the closest multiple of
the base resolution.
Syntax
[:CONFigure]:STEP:CVH{<NRf>[ V ] | MINimum |
MAXimum}
Query Syntax
[:CONFigure]:STEP:CVH?
Parameter
<NRf>[ V ]
Step resolution.
MINIMUM or MIN
Minimum step resolution
MAXIMUM or MAX
Maximum step resolution
Return parameter
<ASCII string>
Returns the range and the
step resolution.
Example
:STEP:CVH 0.01V
Sets the step resolution to 0.01V.
Query example
:STEP:CVH?
> CVH:0.01
Returns the step resolution (0.01V).
[:CONFigure]:STEP:CVL
Set
Query
Description
Configures the step resolution for CV Low Range.
Note: The step resolution setting will be
automatically rounded to the closest multiple of
the base resolution.
Syntax
[:CONFigure]:STEP:CVL{<NRf>[ V ] | MINimum |
MAXimum}
Query Syntax
[:CONFigure]:STEP:CVL?
Parameter
<NRf>[ V ]
Step resolution.
MINIMUM or MIN
Minimum step resolution
MAXIMUM or MAX
Maximum step resolution
Return parameter
<ASCII string>
Returns the range and the
step resolution.